Sample gap register
A sample of the register and risk matrix delivered with a Data Integrity Assessment, so you can see the format and depth before you ask for anything.
This is a fictional example. All laboratories, systems, people and findings on this page are invented for illustration. They do not describe any real company or laboratory.
The register
Every finding is recorded with the system it applies to, the requirement it relates to, a rating from the risk matrix below, and a recommended action. A real register is delivered as a working spreadsheet your team can track remediation in.
| ID | System | Finding | Reference | Impact / likelihood | Risk | Recommended action |
|---|---|---|---|---|---|---|
| DI-001 | FTIR spectrometer workstation | A single shared analyst account is used, so results cannot be attributed to an individual. | 21 CFR 11.10(d), 11.100(a); ALCOA++ Attributable | High / High | High | Create individual accounts. Until then, keep a contemporaneous equipment use log with second-person verification. |
| DI-002 | HPLC data system | The audit trail is enabled but is not part of routine data review before results are released. | FDA Data Integrity Guidance (2018); 21 CFR 11.10(e) | High / Medium | High | Add audit trail review to the data review procedure, using a report filtered to the events that matter. |
| DI-003 | Chromatography workstation | Analysts hold local administrator rights, so raw data files can be deleted or changed outside the application. | 21 CFR 11.10(d), 11.10(g) | High / Medium | High | Move analysts to standard accounts. Hold administrator access with IT, with every use logged and justified. |
| DI-004 | Moisture analyzer | Thermal printouts are treated as the original record, and no procedure defines the record of truth. | 21 CFR 211.180, 211.194(a); ALCOA++ Original, Enduring | Medium / Medium | Medium | Define the record of truth in a procedure. Capture results electronically, or create verified true copies on durable media. |
| DI-005 | HPLC data server | Backups run nightly, but a restore has never been tested. | 21 CFR 11.10(c), 211.68(b); ALCOA++ Enduring, Available | High / Low | Medium | Test a restore on a defined schedule, and document that the restored data is complete and readable. |
| DI-006 | Karl Fischer titrator | Results recorded in a paper logbook are not reconciled with the electronic data. | 21 CFR 211.194(a); ALCOA++ Complete, Consistent | Medium / Medium | Medium | Define the record of truth, and add a reconciliation check to second-person review. |
| DI-007 | Titrator workstation | The workstation runs an unsupported operating system that can no longer be patched. | 21 CFR 211.68(b); ALCOA++ Enduring | Medium / Medium | Medium | Isolate the workstation on the network and restrict removable media. Evaluate virtualization or replacement through change control. |
| DI-008 | Analytical balance | Users can change the system clock, so timestamps are not protected. | 21 CFR 11.10(e); ALCOA++ Contemporaneous | Medium / Low | Low | Restrict clock changes to an administrator, and verify time settings during periodic review. |
| DI-009 | Equipment use log | Entries are sometimes completed after the work is finished. | FDA Data Integrity Guidance (2018); ALCOA++ Contemporaneous, Attributable | Low / Medium | Low | Move to a digital equipment use log completed at the time of use, and reinforce expectations in training. |
| DI-010 | Training records | Records show that the data integrity procedure was read, but not that analysts can perform the instrument-specific checks. | 21 CFR 211.25(a); ALCOA++ Consistent | Medium / Low | Low | Add documented on-the-job training for the instrument-specific data integrity checks. |
The risk matrix
Each finding is placed by its impact on product quality and data reliability, and by the likelihood that it will occur. The resulting rating sets the order of remediation.
- High: act first, with interim controls in place while longer fixes land.
- Medium: plan and schedule in the short term.
- Low: address as part of routine improvement.
| Likelihood: low | Likelihood: medium | Likelihood: high | |
|---|---|---|---|
| Impact: high | MediumDI-005 | HighDI-002, DI-003 | HighDI-001 |
| Impact: medium | LowDI-008, DI-010 | MediumDI-004, DI-006, DI-007 | HighNone |
| Impact: low | LowNone | LowDI-009 | MediumNone |
Want a register like this for your systems?
A free, one-hour scoping call is enough to define the assessment and give you a fixed quote.